--. 0

Product

 [XM8171]  Pipeline CO2 sensors

XM8171

Technical Parameters

Technical parameterParameter value
BrandXUNCHIP
CO2 range0~5000ppm
CO2 accuracy±100ppm
InterfaceRS485/4-20mA/DC0-10V
PowerDC5~24V 1A
PowerDC12~24V 1A
Running temperature-30~85℃
Working humidity5%RH~90%RH

Product Selection

    Product DesignRS485,4-20mA,DC0-10VMultiple output methods, the products are divided into the following models depending on the output method.
Product model output method
XM8171BRS485总线
XM8171M4-20mA
XM8171V10DC0-10V

Product Size

XM8171 XM8171 XM8171 XM8171 XM8171 XM8171

Why choose this product?

XM8171 XM8171 XM8171 XM8171

How to use?

XM8171

Product List

XM8171

Pictures

XM8171 XM8171 XM8171

Communication Protocol

    The product uses RS485 MODBUS-RTU standard protocol format, all operation or reply commands are hexadecimal data. The default device address is 1 when the device leaves the factory, and the module or NON-Recorder default baud rate is 9600,8,n,1 ,but data recorder default baud rate is 115200 .
1. Read data (function code 0x03)
    Inquiry frame (hexadecimal), sending example: query 1 data of 1# device, the upper computer sends the command: 01 03 00 00 00 01 84 0A .
Address Function Code Start Address Data Length Check Code
010300 0000 0184 0A
    For the correct query frame, the device will respond with data: 01 03 02 00 79 79 A6 , response format:
Address Function Code LengthData 1Check Code
0103 0200 79 79 A6
    Data description: The data in the command is hexadecimal, take data 1 as an example, 00 79 is converted to decimal value as 121, assuming the data magnification is 100, then the real value is 121/100=1.21, Others and so on.
2. Common data address table
Configuration Address Register Address Register Description Data Type Value Range
4000100 00carbon dioxideRead Only0~65535
4010100 64Model CodeRead/Write0~65535
4010200 65 total number of measuring points read/write 1~20
4010300 66device addressread/write1~249
4010400 67baud rateread/write0~6
4010500 68communication moderead/write1~4
4010600 69protocol typeread/write1~10

  For example, the range is 400~5000ppm, the analog output is 4~20mA current signal, carbon dioxide and current The calculation relationship is as shown in the formula: C = (A2-A1) * (X-B1) / (B2-B1) + A1, where A2 is carbon dioxide range upper limit, A1 is the lower limit of the range, B2 is current output range upper limit, B1 is the lower limit, X is the currently read carbon dioxide value, and C is the calculated current value. The list of commonly used values is as follows:
current(mA)carbon dioxideValue (ppm) Calculation Process
4400.0(5000-400)*(4-4)÷(20-4)+400
5687.5(5000-400)*(5-4)÷(20-4)+400
6975.0(5000-400)*(6-4)÷(20-4)+400
71262.5(5000-400)*(7-4)÷(20-4)+400
81550.0(5000-400)*(8-4)÷(20-4)+400
91837.5(5000-400)*(9-4)÷(20-4)+400
102125.0(5000-400)*(10-4)÷(20-4)+400
112412.5(5000-400)*(11-4)÷(20-4)+400
122700.0(5000-400)*(12-4)÷(20-4)+400
132987.5(5000-400)*(13-4)÷(20-4)+400
143275.0(5000-400)*(14-4)÷(20-4)+400
153562.5(5000-400)*(15-4)÷(20-4)+400
163850.0(5000-400)*(16-4)÷(20-4)+400
174137.5(5000-400)*(17-4)÷(20-4)+400
184425.0(5000-400)*(18-4)÷(20-4)+400
194712.5(5000-400)*(19-4)÷(20-4)+400
205000.0(5000-400)*(20-4)÷(20-4)+400
    As shown in the above formula, when measuring 8mA, current current is 1154ppm。
  For example, the range is 400~5000ppm, the analog output is 0~10V DC0-10Vvoltage signal, carbon dioxide and DC0-10Vvoltage The calculation relationship is as shown in the formula: C = (A2-A1) * (X-B1) / (B2-B1) + A1, where A2 is carbon dioxide range upper limit, A1 is the lower limit of the range, B2 is DC0-10Vvoltage output range upper limit, B1 is the lower limit, X is the currently read carbon dioxide value, and C is the calculated DC0-10Vvoltage value. The list of commonly used values is as follows:
DC0-10Vvoltage(V)carbon dioxideValue (ppm) Calculation Process
0400.0(5000-400)*(0-0)÷(10-0)+400
1860.0(5000-400)*(1-0)÷(10-0)+400
21320.0(5000-400)*(2-0)÷(10-0)+400
31780.0(5000-400)*(3-0)÷(10-0)+400
42240.0(5000-400)*(4-0)÷(10-0)+400
52700.0(5000-400)*(5-0)÷(10-0)+400
63160.0(5000-400)*(6-0)÷(10-0)+400
73620.0(5000-400)*(7-0)÷(10-0)+400
84080.0(5000-400)*(8-0)÷(10-0)+400
94540.0(5000-400)*(9-0)÷(10-0)+400
105000.0(5000-400)*(10-0)÷(10-0)+400
    As shown in the above formula, when measuring 5V, current DC0-10Vvoltage is 2300ppm。
software for test product
After downloading, please read the readme.txt documentation

Data and Software
software for test product
Data and Software
software for test product
After downloading, please read the readme.txt documentation

Data and Software

 Online shop 

 Contact Online 

Email: sale@sonbest.com

TYPE: XM8171

WeChat: 13918126020

SKYPE: SOOBUU

Phone: +86-021-66862075

QQ1: 71182854 QQ2:613035997